NEULOG CALCIUM ION SENSOR GUIDE
NeuLog calcium sensor
NUL-238 Carolina # 369667
The NeuLog calcium ion sensor can be used for any science experiment or activity
which can utilize calcium concentration readings such as in the fields of Environmental
Science, Ecology, Chemistry, Biology, etc.
Just a few of the thousands of possible experimental subjects that can be done with
the NUL-238 sensors are: chemical reactions, water quality, environmental health,
small animal behavior, ecological studies, and many more
The calcium sensor uses the following units of measure:
Milligrams per liter (mg/L): SI unit of measure for density of a solid dissolved
in a liquid.
Parts per million (ppm): The total amount of a specific molecule (calcium for
example) per million molecules in a sample.
Calcium sensor usage:
The calcium sensor’s electrode uses a PermaFil (non-refillable), Ion-selective
electrode. The reference chamber is gel filled and sealed off and therefore no further
reference fill solution is required.
Usage guide:
Concentration range
1M to 5x10
-7
M
0.02 to 40,000 ppm
pH range
2.5 to 11 pH
Temperature range
0 to 40°C
Minimum sample size
3mL in a 50 mL beaker
Reproducibility
±4%
Electrode resistance
1 to 4 megaohms
Required equipment:
Wash bottle with distilled or deionized water.
Several clean beakers.
1 mL, 10 mL, and 100 mL pipettes.
Technical notes:
All standard solutions should be prepared fresh.
The standards should be at the same temperature as the sample.
Use ISA in all solutions.
Note: ISA is added to all solutions to ensure the samples and standards
have the same ionic strength. The ratio is 1:50
Electrode preparation:
1. Remove the protective cap encasing the calcium probe’s electrode.
CAUTION: Do not touch the PVC membrane at the end of the probe
with your fingers
2. Rinse the electrode with deionized water and blot dry. Be sure not to rub as
this will harm the sensor.
3. Soak the electrode in deionized water for 10 minutes, then in a diluted
calcium standard solution (0.0001 M / 1 ppm) until ready for use.
Calibration:
Be sure to calibrate the calcium sensor once daily during periods of regular usage.
1. Prepare the electrode as described in the “Electrode preparation” guide
above.
2. Place 97 mL of deionized water into a 150 mL beaker.
3. Add 2 mL of Ionic Strength Adjustor (ISA) (comes with sensor).
4. Rinse the electrode with deionized water and blot dry and place in the 150
mL beaker.
5. Pipette 1 mL of 1000 ppm calcium standard solution (comes with sensor)
and stir thoroughly.
6. In the NeuLog software, once the sensor has been detected, click on the
calcium sensor icon.
7. After a stable reading has been reached, click on the calibrate button, or
hold the “Start/Stop” button located on the sensors faceplate down for 4
seconds.
8. Rinse electrode once more with deionized water and blot dry.
9. The calcium sensor is now calibrated and ready for use.
Vue de la page 0
1 2 3 4 5

Résumé du contenu

Page 1

NEULOG CALCIUM ION SENSOR GUIDE NeuLog calcium sensor NUL-238 Carolina # 369667 The NeuLog calcium ion sensor can be used for any science experim

Page 2

NEULOG CALCIUM ION SENSOR GUIDE Electrode storage: Short term (overnight or weekend): Rinse the electrode thoroughly with deionized water and pla

Page 3

NEULOG CALCIUM ION SENSOR GUIDE 5. The WIFI-201 indicator lights will flash; take no action until the LED to the far left turns blue. 6. Take your

Page 4

NEULOG CALCIUM ION SENSOR GUIDE Restoring sensor’s factory default settings: Procedure: 1. Connect the USB-200 to a PC, Mac, tablet, or smart dev

Page 5

NEULOG CALCIUM ION SENSOR GUIDE Maintenance and storage: Before use, please review the “Calcium sensor usage”, “Electrode preparation”, “E