NEULOG OXYGEN SENSOR GUIDE
NeuLog oxygen sensor
NUL-205 Carolina # 369634
The oxygen sensor can be used for any science experiment or activity where oxygen
levels, dissolved or gaseous, are required. Some fields of study include Chemistry,
Physiology, Exercise Science, Biochemistry, Biology, Earth Science, Environmental
Science, etc.
Just a few of the thousands of possible experiments that can be done with the NUL-
205 sensor are: Monitoring of ecological systems, combustion reactions, oxygen
consumption rates, studying photosynthesis, testing chemical solutions, hydrogen
peroxide decomposition, and many more.
This sensor is practical for testing both in school laboratories and in the field and has
two modes of operation: open air mode to test gaseous oxygen concentration
(percentage of oxygen in the air), and dissolved oxygen mode for the testing of liquids
(mg/L and percentage dissolved oxygen in a liquid)
Units of measure this sensor measures are:
% oxygen concentration in air: The concentration of gaseous oxygen
present in the sample.
Mg/L: Total mass of dissolved oxygen per liter of a given liquid sample.
% oxygen in water: The percentage of dissolved oxygen present out of an
entire liquid sample.
Sensor usage and storage:
Prior to use:
1. Remove the blue rubber cap from the NeuLog oxygen probe.
2. Unscrew the tip-end membrane cap of the probe, inside you will see a metal
electrode.
3. Add 1mL of the “DO Filling Solution” directly into the tip-end which was just
unscrewed.
4. Carefully reattach the tip-end of the probe by screwing it back into the
probe.
5. Do not replace the blue rubber cap until after testing.
Storage:
Short-term (less than 24 hours):
For short term storage, fill a beaker with enough distilled water to cover the membrane
cap of the oxygen probe. With the membrane cap still attached to the probe, place it
tip side down in the distilled water. This easy storage method is ideal for short term
storage, between trials or up to 24 hours.
Long-term (more than 24 hours):
1. Unscrew the membrane cap from the oxygen probe.
2. Discard the DO filling solution from the membrane cap.
3. Rinse the cap with distilled water and gently shake to air dry.
4. Rinse and dry the exposed metal electrode with distilled water; it can be
dried with lab wipes.
5. For storage purposes: Loosely reattach the membrane cap to the oxygen
probe.
Sensor calibration:
The NeuLog oxygen sensor does not need to be calibrated prior to each use because
the prior calibration settings are stored in the sensors software. We recommend
recalibrating prior to discrete data collection such as reading the specific oxygen
content of water samples.
Open air mode:
The oxygen sensor needs to be calibrated before use. Calibration of the probe
assumes a standard atmospheric oxygen concentration of 20.9% (assuming stable
levels in the Earth‟s atmosphere at sea level). The simplest calibration is to hold the
button on the oxygen sensor down for 3-4 seconds.
Vue de la page 0
1 2 3 4 5 6

Résumé du contenu

Page 1 - NEULOG OXYGEN SENSOR GUIDE

NEULOG OXYGEN SENSOR GUIDE NeuLog oxygen sensor NUL-205 Carolina # 369634 The oxygen sensor can be used for any science experiment or activity wh

Page 2

NEULOG OXYGEN SENSOR GUIDE Materials Needed:  NUL-205 oxygen sensor  1 of the following (depending on experiment type) o USB-200 USB Module (

Page 3

NEULOG OXYGEN SENSOR GUIDE Procedure: 1. Install the Neulog software 2. Connect the USB-200 Module to the PC or Mac 3. Connect the oxygen Sensor

Page 4

NEULOG OXYGEN SENSOR GUIDE Operation with NeuLog viewer Prior to initial use the Neulog oxygen sensor needs to be calibrated, please review the “Se

Page 5

NEULOG OXYGEN SENSOR GUIDE Included with sensor:  NeuLog NUL-205 Oxygen Sensor Guide (this document).  Neulog oxygen sensor probe which is atta

Page 6

NEULOG OXYGEN SENSOR GUIDE Warranty: We promise to deliver our sensor free of defects in materials and workmanship for a period of 3 years from the