NEULOG CHLORIDE SENSOR GUIDE
Electrode storage:
Short term (overnight or weekend):
Rinse the electrode thoroughly with deionized water and place the tip in a diluted
standard solution (around 10 ppm) between measurements
Long term:
Rinse the electrode thoroughly with deionized water and blot dry. Attach the hard
plastic protective electrode cap again to protect the sensing element.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-239 Chloride Sensor
USB-200 USB Module
A USB to mini USB cable (which comes with the USB-200)
Before use, please review the “Chloride sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Calibration” sections of this document above to ensure the
proper use and longevity of the NeuLog chloride ion sensor.
Your chloride sensor needs to be connected to a USB-200 module. The USB-200
module then connects to a computer via a USB to mini-USB cable. Please note that
you cannot plug the chloride sensor directly into the computer.
Resident PC software and browser based software can be downloaded for free at
www.NeuLog.com/dowload as well as a full software user guide. Note: Make sure not
to download and install both types of software, they will conflict on the computer.
Procedure:
1. Install the NeuLog software
2. Connect the USB-200 module to the PC or Mac
3. Connect the chloride sensor to the USB-200 module (they directly plug
together).
4. Open the NeuLog software.
5. Once a chloride sensor logo appears on the left side of the screen the probe
has been automatically identified and you can begin experimentation.
6. If the chloride sensor is not automatically identified then click the “Search for
sensors” icon to find the sensor.
7. Select the “On-line experiment” button; this will open a graph below.
8. Click on the “Module setup” button located on the chloride sensor icon in the
module window to change the sensor settings if need be.
9. Click on the experiment set up button to change the experiment settings if
need be (experiment duration for example).
10. The chloride sensor will give a live reading in the box to the left of the
screen while plugged in.
11. To run an experiment and collect data click “Run experiment”.
12. To end data collection early, click “Stop experiment”.
Tablet, smart phone device
Materials needed:
NUL-239 Chloride Sensor
WIFI-201 WIFI module
BAT-200 Battery
Before use, please review the “Chloride sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Calibration” sections of this document above to ensure the
proper use and longevity of the NeuLog chloride ion sensor.
Your chloride sensor needs to be connected to a WIFI-201 module. The WIFI-201
module will create a closed NeuLog wifi network which will stream the NeuLog data to
a device of your choosing. Once your device is wirelessly connected to the NeuLog
network you can run experiments and collect data through a browser of your choosing.
Procedure:
1. Connect the chloride sensor directly to the left side of a WIFI-201 module
(no wires required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet
using a USB to micro USB charger (such as a typical cell phone charger).
The WIFI-201 module will run for 15-75 minutes (depending on the sensor)
without being plugged in.